HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00563
- Event Type
- Injury
- Date Received
- October 8, 2020
- Report Date
- October 5, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THERE WAS NO KNOWN PATIENT INVOLVEMENT. SERIAL NUMBER IS UNKNOWN. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA WILL ATTEMPT TO RETRIEVE ADDITIONAL DETAILS. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THROUGH LITERATURE REVIEW LIVANOVA BECAME AWARE OF THREE (3) PATIENTS INFECTED WITH MYCOBACTERIUM CHIMAERA IN (B)(6). THE THREE PATIENTS DIAGNOSED IN 2019 WERE DETECTED BY PASSIVE CASE FINDING DUE TO THE CLINICAL AWARENESS OF THEIR CLINICIANS. WITHIN THE ARTICLE IT IS STATED THAT THESE EVENTS HAVE BEEN REPORTED TO THE HEALTH AUTHORITIES. THE HOSPITAL/HOSPITALS WERE THE SURGERIES TOOK PLACE AND THE DATE OF THE SURGERIES ARE UNKNOWN, AS WELL AS THE SERIAL NUMBER OF THE HEATER-COOLER USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1114519 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |