CARDIOMEMS PATIENT ELECTRONIC SYSTEM
Report
- Report Number
- 3004936110-2020-00613
- Event Type
- Malfunction
- Date Received
- October 8, 2020
- Date of Event
- September 24, 2020
- Report Date
- January 29, 2021
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- MOM
- UDI-DI
- 05414734509800
- PMA / PMN Number
- P100045
- Removal / Correction Number
- 3004936110-06/01/18-001R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
Narratives
A FINDING OF INCORRECT SPEED SETTING ON COMPACT FLASH WAS FOUND AND THE ROOT CAUSE WAS CONCLUDED TO BE RC064-01 COMPACT FLASH-ERROR 5. A REVIEW OF THE DHR WAS PERFORMED FOR BATCH #6243387 AND RULED OUT THE POSSIBILITY OF A MANUFACTURING RELATED ISSUE CAUSING OR CONTRIBUTING TO THE REPORTED COMPLAINT ISSUE, AS NO NON-CONFORMANCE RELATED TO THE REPORTED ISSUE WAS FOUND IN THE BATCH RECORDS REVIEWED. THE PATIENT ELECTRONIC UNIT RECEIVING AN ERROR 5 MESSAGE WAS CONFIRMED AND THE ROOT CAUSE WAS CONCLUDED TO BE COMPACT FLASH-ERROR 5.
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. A REPLACEMENT UNIT RESOLVED THE ISSUE.
THE DEVICE IS INCLUDED IN THE ERROR 5 ADVISORY NOTICE ISSUED BY ABBOTT ON 01 JUNE 2018. THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN THE FINAL REPORT.
IT WAS REPORTED THAT THE PATIENT ELECTRONIC UNIT RECEIVED AN ERROR 5 MESSAGE, WHICH INDICATES AN ISSUE RELATED TO TEMPERATURE AND BAROMETRIC PRESSURE. UPDATING THE UNIT VIA MERLIN.NET RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1111896 | CARDIOMEMS PATIENT ELECTRONIC SYSTEM | SYSTEM, HEMODYNAMIC, IMPLANTABLE | MOM | ST. JUDE MEDICAL, INC. | CM1100 | 6243387 | 05414734509800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |