FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 31GA 6MM WHOLEUNIT 10BAG

MDR report key: 10646288 · Received October 7, 2020

Report

Report Number
1920898-2020-01375
Event Type
Malfunction
Date Received
October 7, 2020
Date of Event
September 17, 2020
Report Date
October 30, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00382903249091
PMA / PMN Number
K024112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 10/23/2020. H.6. INVESTIGATION: CUSTOMER RETURNED (4) LOOSE 3/10CC SYRINGES. CUSTOMER STATES THAT THE NEEDLE HUB SEPARATED WHEN REMOVING SHIELD, THE SHIELDS ARE DIFFICULT TO REMOVE, AND THEY ARE TIGHT. ALL RETURNED SYRINGES WERE EXAMINED AND ALL EXHIBITED THE HUB-NEEDLE/SHIELD ASSEMBLY SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0083503 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. PROCESS SUMMARY: AUTOMATIC SYRINGE ASSEMBLY MACHINE, WHICH FEEDS 3/10CC, SYRINGE COMPONENTS (BARREL, STOPPER, PLUNGER, NEEDLE ASSEMBLY & CAP) AND ASSEMBLES THESE COMPONENTS. THIS MACHINE CONSISTS OF A BARREL CLEANING DIAL, LUBRICATION DIAL, PLUNGER/STOPPER ASSEMBLY DIAL, SYRINGE ASSEMBLY DIAL, AND VARIOUS INSPECTIONS AND TRANSFER DIALS. DHR, L2L DISPATCHES, AND LOGBOOK ENTRIES WERE LOOKED AT, NOTHING PERTAINING TO THIS DEFECT WAS FOUND. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1630423 WAS INITIATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.3ML 31GA 6MM WHOLEUNIT 10BAG NEEDLE HUB SEPARATED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 324909 BATCH NO. 0083503. IT WAS REPORTED THAT NEEDLE SHIELDS ARE TIGHT AND THE NEEDLE HUB SEPARATES.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.3ML 31GA 6MM WHOLEUNIT 10BAG NEEDLE HUB SEPARATED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 324909 BATCH NO. 0083503. IT WAS REPORTED THAT NEEDLE SHIELDS ARE TIGHT AND THE NEEDLE HUB SEPARATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1104939 SYRINGE 0.3ML 31GA 6MM WHOLEUNIT 10BAG PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 324909 0083503 00382903249091

Patients

Seq Age Sex Outcome Treatment
1 Other