X-SMART PLUS CONTRA ANGLE
Report
- Report Number
- 8031010-2020-00211
- Event Type
- Malfunction
- Date Received
- October 7, 2020
- Report Date
- October 2, 2020
- Manufacturer
- DENTSPLY MAILLEFER
- Product Code
- EGS
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY, OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. RECEIVED - 1X X-SMART PLUS CONTRA-ANGLE 6:1 A103300000000; SN: (B)(4). 1X X-SMART PLUS HANDPIECE A103400000300; SN: (B)(4). 1X X-SMART PLUS UNIV.AC.ADAPTER A103400000200. 1X X-SMART PLUS UNIT SN: (B)(4). X-SMART PLUS BATTERY. DEFECTIVE BATTERY. BATTERY LIFE IS LOW. X-SMART PLUS HEAD CAP. BAD MAINTENANCE (USER). THERE IS SOME FOREIGN MATTER ON THE HEAD CAP. X-SMART PLUS CARTRIDGE; VARIOUS MECHANICAL PROBLEM; FILES CANNOT LOCK BECAUSE HOOK OF CARTRIDGE IS DAMAGED.
IN THIS EVENT IT WAS REPORTED THAT A X-SMART PLUS CONTRA ANGLE WON'T HOLD FILES; NO INJURY RESULTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1106361 | X-SMART PLUS CONTRA ANGLE | HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL | EGS | DENTSPLY MAILLEFER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |