FDA Adverse Event Malfunction Summary report: N

FLEXIBLE BIOPSY CUP 3FR 115CM

MDR report key: 10644648 · Received October 7, 2020

Report

Report Number
3003790304-2020-00099
Event Type
Malfunction
Date Received
October 7, 2020
Report Date
October 7, 2020
Manufacturer
LAKE REGION MEDICAL
Product Code
FCL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER RETURNED A FBC-3115 (FLEXIBLE BIOPSY CUP 3FR 115CM) WITH LOT # 4923905 FOR EVALUATION. THE UNIT WAS RECEIVED WITHOUT THE ORIGINAL PACKAGING. A VISUAL INSPECTION WAS PERFORMED ON THE RECEIVED DEVICE AND FOUND NO VISUAL EVIDENCE OF DAMAGES ON THE COIL SHEATH, THE CUP WAS IN GOOD CONDITION. THE FORCEP CUPS WERE IN THE CLOSED POSITION WHEN RECEIVED. FOR TESTING, THE THUMB RING WAS ACTIVATED AND THE CUP WAS UNABLE TO OPEN WHICH CONFIRMED THE USER¿S COMPLAINT. FURTHER EVALUATION ON THE MOVEMENT OF THE THUMB RING, OBSERVED THAT THE COIL SPRING AT THE DISTAL END OF THE HANDLE BECAME LOOSE DURING ACTIVATION WHICH RESULTED IN THE FAILURE MODE OF THE CUPS OPERATION. BASED ON THE EVALUATION, THE UNIT IS UN-OPERABLE DUE TO THE DAMAGE TO COIL SHEATH SECTION. THE INSTRUCTION MANUAL ON PAGE 5 IN THE ¿OPERATIONAL TEST¿ SECTION BULLET 2 STATES ¿USING THE HANDLE AND THUMB RING, OPEN AND CLOSE THE FORCEPS A FEW TIMES TO VERIFY SMOOTH OPERATION. ALWAYS OPERATE THE FORCEPS LIGHTLY, AS EXCESSIVE FORCE ON THUMB RING (EITHER PUSHING OR PULLING) CAN CAUSE DETERIORATION AND BREAKAGE¿.

Description of Event or Problem · 1

THE SERVICE CENTER WAS INFORMED THAT DURING PREPARATION FOR USE, THE CUPS OF TWO FORCEPS WERE NO LONGER OPENING. THE FACILITY REPORTED THAT THE FORCEPS WERE USED LESS THAN A YEAR. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS IS 1 OF 2 REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1107738 FLEXIBLE BIOPSY CUP 3FR 115CM FLEXIBLE BIOPSY CUP 3FR 115CM FCL LAKE REGION MEDICAL FBC-3115 4923905

Patients

Seq Age Sex Outcome Treatment
1