FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10641555 · Received October 7, 2020

Report

Report Number
9611109-2020-00561
Event Type
Malfunction
Date Received
October 7, 2020
Date of Event
September 9, 2020
Report Date
October 5, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE ERROR CODE WAS DISPLAYED ON THE PATIENT SIDE AND THAT WAD DUE TO A DEFECTIVE STIRRER MOTOR. THE MOTOR WAS REPLACED AS WELL AS DISTRIBUTION BOARD AND CPU BOARD. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Additional Manufacturer Narrative · 1

THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-95 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DISPLAYED AN ERROR CODE DURING SETUP. THE SIDE AFFECTED (PATIENT OR CARDIOPLEGIA) IS STILL UNKNOWN. A MALFUNCTION OF THE STIRRER MOTOR CANNOT BE EXCLUDED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1105729 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-95

Patients

Seq Age Sex Outcome Treatment
1