FDA Adverse Event
Malfunction
Summary report: N
DIMENSION CLINICAL CHEMISTY SYSTEM
MDR report key: 1063859
·
Received June 19, 2008
Report
- Report Number
- 1226181-2008-00067
- Event Type
- Malfunction
- Date Received
- June 19, 2008
- Date of Event
- May 30, 2008
- Report Date
- May 30, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC
- Product Code
- MMI
- PMA / PMN Number
- K010061
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REP (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS SLIPPING HM WAS PUMP, DUE TO LACK OF MONTHLY MAINTENANCE BY THE OPERATOR. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED ON THE SAME ANALYZER AND A NEGATIVE RESULT WAS OBTAINED. PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC | XPANDPLUS HM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |