DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00065
- Event Type
- Malfunction
- Date Received
- June 19, 2008
- Date of Event
- May 24, 2008
- Report Date
- May 24, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K010061
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REP (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS R1 ARM AND DRAIN PROBLEMS CAUSING CARRY INTO THE CUVETTE. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
TWO FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON A PT SAMPLE. THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED ON THE SAME ANALYZER AND A NEGATIVE RESULT WAS OBTAINED. PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | XPANDHM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR |