FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PICE FOLDABLE INTRAOCULA

MDR report key: 1063754 · Received June 19, 2008

Report

Report Number
2023826-2008-00824
Event Type
Malfunction
Date Received
June 19, 2008
Date of Event
April 30, 2008
Report Date
May 22, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION RESULTS - (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT A HAPTIC PLATE WAS TORN AND A PIECE OF IT WAS TORN OFF AND MISSING. THERE WAS EVIDENCE OF A CLEAR SURGICAL RESIDUE. CONCLUSION: (OTHER) - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGES DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT HAD A CC4204BF COLLAMER PLATE LENS, AND THE LENS WAS SCRATCHED WHILE ADVANCING IN THE INJECTOR. THERE WAS PT CONTACT BUT NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PICE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CC4204BF NA

Patients

Seq Age Sex Outcome Treatment
1 UNK INJECTOR MODEL NANOPOINT| CARTRIDGE MODEL UNK