FDA Adverse Event Malfunction Summary report: N

ISITE PACS

MDR report key: 1063669 · Received June 20, 2008

Report

Report Number
2954704-2008-00004
Event Type
Malfunction
Date Received
June 20, 2008
Date of Event
May 22, 2008
Report Date
June 18, 2008
Manufacturer
PHILIPS HEALTHCARE INFORMATICS, INC.
Product Code
LLZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE FROM RESERVE SAMPLE EVALUATED. THE MPR FEATURE IS ONLY IN 3.5.X VERSIONS OF ISITE AND IS NOT UTILIZED BY THE MAJORITY OF USER FACILITIES. DUE TO THIS ISSUE AND PREVIOUSLY IDENTIFIED DEFECTS, A PRODUCT NOTIFICATION WILL BE SENT OUT TO ALL 3.5 CUSTOMERS AND THE FEATURE WILL BE DISABLED FROM ALL ISITE 3.5.X INSTALLS.

Description of Event or Problem · 1

THE SCOUT LINES AND LOCALIZER CROSSHAIRS WHEN ENABLED ON MULTI PLANAR RECONSTRUCTION (MPR) IMAGES, CREATED BY ISITE UNDER THE FOLLOWING CONDITIONS, WILL DISPLAY AT THE WRONG TRIANGULATION POSITION REPRESENTING THE IMAGE LOCATIONS OF THE OTHER ORTHOGONAL PLANES. THE IMAGE DATA IS ACQUIRED WHEN GANTRY IS TILTED. ISITE SYSTEM PREFERENCE UNDER "MAXIMUM VARIANCE FOR LINKING" IS CONFIGURED WITH MORE THAN 2 DEGREES FROM THE ANGLE OF THE GANTRY TILT DURING ACQUISITION OF THE ORIGINAL DATASET. MPR IMAGES ARE NOT AFFECTED BY THIS ERROR - ONLY THE SCOUT LINE AND LOCALIZER CROSSHAIR PLACEMENT ON THE MPR IMAGES ARE AFFECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISITE PACS PICTURE ARCHIVING AND COMMUNICATION SYS LLZ PHILIPS HEALTHCARE INFORMATICS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other