FDA Adverse Event Malfunction Summary report: N

COBAS 6000 C501 MODULE

MDR report key: 1063644 · Received June 19, 2008

Report

Report Number
1823260-2008-04851
Event Type
Malfunction
Date Received
June 19, 2008
Date of Event
May 27, 2008
Report Date
June 19, 2008
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CEM
PMA / PMN Number
K060373
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

ONE PATIENT WITH DISCREPANT CALCIUM RESULTS. INITIAL RESULT 0.3 MG/DL, REPEAT 9.3 MG/DL. THE INITIAL RESULT WAS NOT REPORTED. THE FIELD SERVICE REPRESENTATIVE WAS UNABLE TO DETERMINE ROOT CAUSE BUT NOTED HE CLEANED THE PROBE AND CHANGED OUT THE VALVE ASSEMBLY BEHIND THE SYRINGES. PERFORMANCE TESTS PERFORMED WHICH WERE WITHIN SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS 6000 C501 MODULE CLINICAL CHEMISTRY ANALYZER - CEM CEM ROCHE DIAGNOSTICS C501

Patients

Seq Age Sex Outcome Treatment
1 UNK