FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK MR BALLOON CATHETER

MDR report key: 1063640 · Received June 19, 2008

Report

Report Number
2134265-2008-01737
Event Type
Malfunction
Date Received
June 19, 2008
Date of Event
April 1, 2008
Report Date
May 20, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
LOX
PMA / PMN Number
P860019/S167
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RETURNED PRODUCT ANALYSIS CONFIRMED THE DIFFICULTY STATED IN THE COMPLAINT. THE BALLOON CATHETER WAS RETURNED IN A DEFLATED STATE. BLOOD WAS OBSERVED IN THE BALLOON AND INFLATION LUMEN. MICROSCOPIC INSPECTION FOUND A PINHOLE LOCATED OVER THE PROXIMAL MARKERBAND. MICROSCOPIC EXAMINATION OF THE AREA SURROUNDING THE PINHOLE DID NOT REVEAL ANY IRREGULARITIES IN THE BALLOON MATERIAL WHICH WOULD HAVE CONTRIBUTED TO THE PINHOLE. THE MOST PROBABLE ROOT CAUSE IS DETERMINED TO BE OPERATIONAL CONTEXT, DUE TO THE LIKELIHOOD OF ANATOMICAL AND OR PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, WHICH CONTRIBUTED TO THE REPORTED EVENT. IT SHOULD BE NOTED THAT THERE WAS NO EVIDENCE OF MANUFACTURING DEFECT OR ANOMALY WITHIN THE MANUFACTURING RECORDS REVIEWED FOR THE REPORTED BATCH.

Description of Event or Problem · 1

EVENT BECAME REPORTABLE BASED UPON ANALYSIS COMPLETED 05/20/2008. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, A BALLOON LEAK OCCURRED. THE SERIOUSLY CALCIFIED 80% STENOTIC LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. THE LESION WAS PRE-DILATED WITH A 2.5 X 15MM UNSPECIFIED BALLOON. AFTER THE PHYSICIAN DEPLOYED AN UNSPECIFIED STENT, THE 3.25 X 15MM QUANTUM MAVERICK BALLOON WAS USED FOR POST-DILATION; AT 18 ATMOSPHERES OF PRESSURE THE BALLOON LEAKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PT INJURIES OR COMPLICATIONS WERE REPORTED. THE PATIENT'S CONDITION WAS REPORTED AS "STABLE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK MR BALLOON CATHETER LOX CATHETER, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC 15 X 3.25 MM 11404754

Patients

Seq Age Sex Outcome Treatment
1 68 YR