FDA Adverse Event Malfunction Summary report: N

AXSYM RUBELLA IGG ANTIBODY

MDR report key: 1063616 · Received June 19, 2008

Report

Report Number
1415939-2008-00142
Event Type
Malfunction
Date Received
June 19, 2008
Date of Event
March 17, 2008
Report Date
June 2, 2008
Manufacturer
ABBOTT LABORATORIES
Product Code
LFX
PMA / PMN Number
K954045
Removal / Correction Number
1415939-6/06/08-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THE RUBELLA CALIBRATION FAILED ON THE AXSYM ANALYZER WITH ERROR CODE 1048: RUBELLA G CALIBRATION FAILURE, MCAL 1 RESULTS TOO HIGH. THE ABBOTT CUSTOMER SERVICE TECHNICIAN (CTA) RECOMMENDED DECONTAMINATING THE THREE LINES AND REPLACING THE SAMPLE PROBE. THE CTA SENT A STANDARD CALIBRATOR. AFTER RECEIPT OF THE STANDARD CALIBRATOR, THE CALIBRATION PASSED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM RUBELLA IGG ANTIBODY MEIA FOR THE MEASUREMENT OF IGG ANTIBODY TO THE RUBELLA VIRUS LFX ABBOTT LABORATORIES NA 56659M201

Patients

Seq Age Sex Outcome Treatment
1 NI AXSYM ANALYZER| RUBELLA MCAL LIST# 3B23-30