AXSYM RUBELLA IGG ANTIBODY
Report
- Report Number
- 1415939-2008-00158
- Event Type
- Malfunction
- Date Received
- June 19, 2008
- Date of Event
- May 14, 2008
- Report Date
- June 2, 2008
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- LFX
- PMA / PMN Number
- K954045
- Removal / Correction Number
- 1415939-06/06/08-0004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- UNKNOWN
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATED THE RUBELLA CALIBRATION FAILED WITH ERROR CODE 1048: CALIBRATION CHECK FAILURE, CAL A/B RATIO TOO HIGH, ON THE AXSYM ANALYZER. THE ABBOTT CUSTOMER TECHNICAL ADVOCATE (CTA) ASKED THE CUSTOMER TO REPLACE THE MATRIX CELL LOT, DECONTAMINATE THE MUP WITH NEW REAGENT, AND RECALIBRATE. THE CALIBRATION FAILED AGAIN. THE CTA ASKED THE CUSTOMER TO REPLACE THE PROCESS PROBE AND RECALIBRATE WHICH AGAIN RESULTED IN A FAILURE ERROR. THE CTA SENT THE CUSTOMER REPLACEMENT CALIBRATOR AND ASKED THE CUSTOMER TO CENTRIFUGE THE CALIBRATOR PRIOR TO USE. CENTRIFUGING RESULTED IN AN ACCEPTABLE CALIBRATION. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AXSYM RUBELLA IGG ANTIBODY | MEIA FOR THE MEASUREMENT OF IGG ANTIBODY TO THE RUBELLA VIRUS | LFX | ABBOTT LABORATORIES | NA | 58295M101 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | RUBELLA MCAL LIST# 3B23-30| AXSYM ANALYZER |