FDA Adverse Event Malfunction Summary report: N

AXSYM RUBELLA IGG ANTIBODY

MDR report key: 1063587 · Received June 19, 2008

Report

Report Number
1415939-2008-00160
Event Type
Malfunction
Date Received
June 19, 2008
Date of Event
May 15, 2008
Report Date
June 2, 2008
Manufacturer
ABBOTT LABORATORIES
Product Code
LFX
PMA / PMN Number
K954045
Removal / Correction Number
1415939-6/06/08-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THE RUBELLA CALIBRATION FAILED WITH ERROR CODE 1111: M-CAL 1 CHECK TOO HIGH, RUBELLA G, ON THE AXSYM ANALYZER. THE ABBOTT CUSTOMER TECHNICAL ADVOCATE (CTA) ASKED THE CUSTOMER TO REPLACE THE PROCESS PROBE, DECONTAMINATE THE LINE 1 TUBING, AND CENTRIFUGE THE CALIBRATOR AND REPEAT THE CALIBRATION. AFTER CENTRIFUGING, THE CALIBRATION PASSED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM RUBELLA IGG ANTIBODY MEIA FOR THE MEASUREMENT OF IGG ANTIBODY TO THE RUBELLA VIRUS LFX ABBOTT LABORATORIES NA 58295M101

Patients

Seq Age Sex Outcome Treatment
1 NA RUBELLA MCAL| AXSYM ANALYZER