ISOTRAK SIM.SY2 STRONTIUM-90 TRANSFER STANDARD

MDR report key: 1062839 · Received June 16, 2008

Report

Report Number
1226007-2008-00001
Event Type
Other
Date Received
June 16, 2008
Date of Event
June 19, 2007
Report Date
June 13, 2008
Manufacturer
QSA-GLOBAL INC.
Product Code
IXD
Removal / Correction Number
1226007-7/30/07-001-C
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

QSA (B) (4)UNDER TOOK CORRECTIVE ACTION IN THE FORM OF RE-LABELING THE DEVICE SHIELD CONTAINER AS REPORTED UNDER 21 USC 360 AS A PRUDENT AND RESPONSIBLE MEASURE TO PREVENT A SIMILAR USER ERROR. AT THE TIME OF THE EVENT INVESTIGATION QSA HAD NO INFO INDICATING THE PT INVOLVED WAS INJURED OR THAT MEDICAL INTERVENTION HAD BEEN NECESSARY. THE POTENTIAL FOR MEDICAL INTERVENTION TO MITIGATE THE RISK OF INJURY TO A PT IN THE EVENT OF A RECURRENCE WAS BASED ON THE LITERATURE PROVIDED BY THE MEDICAL FACILITY CONCERNING THE USE OF THE (B) (6) Y-90 ZEVALIN IN EXCESSIVE DOSE LEVELS. QSA DID NOT AT THE TIME BELIEVE THE CONDITIONS OR REPORTED EVENT MET THE CRITERIA FOR MEDICAL DEVICE REPORTING UNDER 21 CFR 803. THIS WAS BASED ON THE LACK OF EVIDENCE OF INJURY TO THE SUBJECT IN QUESTION OR NEED FOR MEDICAL INTERVENTION. IN ADDITION, QSA WAS AWARE THAT SIGNIFICANTLY LARGER DOSES OF THE (B) (6) HAD BEEN USED WITHOUT ADVERSE EFFECTS. APPARENT CAUSE: THE PRIMARY CAUSE OF THIS INCIDENT IS DEVIATION FROM THE SUPPLIED INSTRUCTIONS FOR USE OF THE SR-90 TRANSFER STANDARD. PROCEEDING WITH THE CALIBRATION PROCESS WITHOUT THE SOURCE CERTIFICATE IS THE PROXIMATE CAUSE OF THE CALIBRATION ERROR LEADING TO THE OVERDOSE. A CONTRIBUTING CAUSE TO THE INCIDENT IS THE AMBIGUOUS LABELING OF THE SOURCE CONTAINER. THE USE OF INCONSISTENT LABELING SET THE STAGE FOR THE ERROR TO OCCUR. COMMON PRACTICE IN THE FIELD IS TO ACCEPT THE INFO ON THE DEVICE OR ITS CONTAINER AS THE CORRECT INFORMATION. CONCLUSION: THE SR-90/Y-90 TRANSFER STANDARD IS LABELED IN A WAY THAT CAN LEAD TO IMPROPER USE OF THE DEVICE WHEN CALIBRATING DOSE CALIBRATORS. AS EVIDENCED BY THE INCIDENT REPORTED, IMPROPER CALIBRATION CAN RESULT IN ADMINISTERING AN OVERDOSE OF THE (B) (6).

Description of Event or Problem · 1

ON (B) (6) 2007, QSA GLOBAL CUSTOMER SERVICE WAS INFORMED BY A (B) (6) OF THE FOLLOWING INCIDENT. (B) (6) PRACTITIONERS IMPROPERLY CALIBRATED DOSE CALIBRATORS USING THE DECAY CORRECTED SR-90 ACTIVITY OF THE SR-90/Y-90 REFERENCE STANDARD INSTEAD OF THE Y-90 EQUIVALENT ACTIVITY AS DESCRIBED IN THE HANDLING INSTRUCTIONS. SINCE THE DECAY CORRECTED SR-90 ACTIVITY WAS LOWER THAN THE Y-90 EQUIVALENT ACTIVITY, SUBSEQUENT DISPENSING MEASUREMENTS OF THE (B) (6) Y-90 (B) (6) WERE IN ERROR. THIS RESULTED IN A PT BEING ADMINISTERED A THERAPEUTIC DOSE IN EXCESS OF THE PRESCRIBED DOSE. THIS INCIDENT OCCURRED (B) (6) 2007. THE INVESTIGATION CONCLUDED THAT THE PRODUCT LABEL ON THE DEVICE CONTAINER MAY HAVE CONTRIBUTED TO THE USER ERROR, AND SHOULD BE AMENDED. THE EFFECT ON THE PT INVOLVED IN THIS INCIDENT IS UNK. THE HOSPITAL CONTACT HAS REFERRED TO THE (B) (6) LABELING PACKAGING INSERT FOR INFO ON POSSIBLE EFFECTS, BUT HAS DECLINED TO PROVIDE FURTHER INFO. NOTE: THIS REPORT IS BEING MADE TO RESOLVE AN OBSERVATION BY AN FDA INSPECTOR AND REPORTED ON FORM (B) (4), DATED 5/23/2008.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISOTRAK SIM.SY2 STRONTIUM-90 TRANSFER STANDARD SOURCE, CALIBRATION, SEALED, NUCLEAR IXD QSA-GLOBAL INC. SIM.SY2 13001B

Patients

Seq Age Sex Outcome Treatment
1 Other