Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED/USED RESERVOIR AND IT FAILED PER SPECIFICATIONS. THE RESERVOIR LEAKED PAST THE FIRST AND SECOND O-RINGS DUE TO PLUNGER DAMAGE WITH SEVERAL INDENTATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED CONTINUOUS PROBLEMS WITH AIR BUBBLES IN THE RESERVOIRS. THE CUSTOMER SPOKE WITH HER DIABETES EDUCATOR ABOUT THE ISSUE, BUT THE PROBLEMS CONTINUE. THE CUSTOMER WAS GIVEN SUGGESTIONS ON HOW TO PREVENT AIR BUBBLES. THE CUSTOMER WILL RETURN A RESERVOIR FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RSVR MMT-326A 10PK PRDGM 1.8ML 13L DISPOSABLES FMF MEDTRONIC MINIMED MMT-326A WA2507144