Additional Manufacturer Narrative · 1
THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD AND THE STEER CASTER WOULD NOT LOCK INTO THE STEER POSITION. THE TECHNICIAN FOUND THE BRAKE-STEER CASTER WAS INSTALLED INCORRECTLY. THE BRAKE/STEER CASTER WAS INSTALLED IN THE CORRECT POSITION TO REPAIR THE BED.
Description of Event or Problem · 1
ALLEGED THE BRAKE PEDAL WILL "POP" OUT OF THE BRAKE POSITION WHEN THE BED IS MOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY 4 | BIRTHING BED | HDD | HILL-ROM RITTER | 3700B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |