FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1062751 · Received June 18, 2008

Report

Report Number
1824206-2008-02699
Event Type
Malfunction
Date Received
June 18, 2008
Date of Event
May 22, 2008
Report Date
May 22, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE/STEER CASTER WAS INSTALLED INCORRECTLY. THE BRAKE/STEER CASTER WAS INSTALLED IN THE CORRECT POSITION TO REPAIR THE BED.

Description of Event or Problem · 1

ALLEGED THE BRAKE WILL NOT LOCK PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM RITTER 3700B

Patients

Seq Age Sex Outcome Treatment
1 UNK