FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 1062748 · Received June 18, 2008

Report

Report Number
1824206-2008-02649
Event Type
Malfunction
Date Received
June 18, 2008
Date of Event
June 2, 2008
Report Date
June 2, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Removal / Correction Number
Z-0240-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CUSTOMER ALLEGED THE LEFT FOOT RAIL IS NOT LATCHING. SHE IS UNAWARE OF ANY INJURIES CAUSED BY THIS ALLEGED NON LATCHING FOOTRAIL. TSR FOUND THAT ONE OF THE RETAINING RINGS (PART NUMBER 3532501) THAT SECURE THE LONG ROD (PART NUMBER 69798) WAS MISSING FROM THE LEFT FOOT RAIL. HE INSTALLED A NEW RETAINING RINGS AND THE SIDERAIL FUNCTIONED AS DESIGNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER P3200 NA

Patients

Seq Age Sex Outcome Treatment
1 NA