FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAL PRDGM INSULIN CL EN

MDR report key: 1062679 · Received June 16, 2008

Report

Report Number
2032227-2008-00988
Event Type
Malfunction
Date Received
June 16, 2008
Date of Event
May 23, 2008
Report Date
April 2, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP GAVE ERROR ALARMS AND FAILED THE DISPLACEMENT TESTS DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT THAT THE INSULIN PUMP WAS GIVING ERROR ALARMS. TROUBLESHOOTING REVEALED THAT THE INSULIN PUMP NEEDED TO BE REPLACED. ADVISED THE CUSTOMER TO REVERT TO A BACK-UP PLAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAL PRDGM INSULIN CL EN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-722LNAL

Patients

Seq Age Sex Outcome Treatment
1