FDA Adverse Event
Injury
Summary report: N
DEXTRUS 4136
MDR report key: 1062671
·
Received June 10, 2008
Report
- Report Number
- 1028232-2008-00595
- Event Type
- Injury
- Date Received
- June 10, 2008
- Date of Event
- March 3, 2008
- Report Date
- May 6, 2008
- Manufacturer
- BIOTRONIK GMBH AND CO.
- Product Code
- DTB
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE, THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR LEAD AND ALSO THE PRODUCTION LOT HAVE BEEN RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST OF THIS LEAD AND THE PRODUCTION LOT. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFO REGARDING A POCKET INFECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4136 | PACER LEADS | DTB | BIOTRONIK GMBH AND CO. | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |