FDA Adverse Event Malfunction Summary report: N

BARD URETERAL CATHETER

MDR report key: 1062661 · Received June 12, 2008

Report

Report Number
1018233-2008-00043
Event Type
Malfunction
Date Received
June 12, 2008
Date of Event
May 15, 2008
Report Date
June 12, 2008
Manufacturer
C.R. BARD, INC.
Product Code
EYB
PMA / PMN Number
PRE-AMEN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE MANUFACTURING LOT NUMBER PROVIDED, PRODUCT IS IN EXCESS OF 34 YEARS OLD. WHILE THE POLYURETHANE DEVICE IS VERY DURABLE, SOME DEGRADATION OF PRODUCT INTEGRITY IS EXPECTED OVER A 34 YEAR PERIOD. A 5 YEAR EXP DATE WAS PLACED ON THESE PRODUCTS IN THE EARLY NINETIES. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF THE SAMPLE EVAL. BASELINE REPORT PREVIOUSLY PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE URETERAL CATHETER BROKE INTO MULTIPLE PIECES DURING A DIAGNOSTIC URETEROSCOPY. THE PIECES OF THE CATHETER THAT ARE IN THE URETER HAVE NOT BEEN REMOVED, AND WILL BE REMOVED IN APPROX 6 WEEKS VIA URETEROSCOPY. THE CATHETER WAS BEING USED DURING AN OPEN PROSTATECTOMY PROCEDURE, SO THAT THE DOCTOR COULD VISUALIZE THE URETER DURING THE PROCEDURE. NO STONES WERE NOTED IN THE PT. THE DOCTOR STATED THAT THE CATHETER BROKE IN THE DISTAL URETER AND PROBABLY BROKE IN ANOTHER LOCATION IN THE URETER. NO IMPACT TO THE PT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BARD URETERAL CATHETER EYB C.R. BARD, INC. NA AL31521

Patients

Seq Age Sex Outcome Treatment
1 UNK