FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 1062653 · Received June 10, 2008

Report

Report Number
1028232-2008-00581
Event Type
Injury
Date Received
June 10, 2008
Date of Event
February 28, 2008
Report Date
May 6, 2008
Manufacturer
BIOTRONIK GMBH AND CO.
Product Code
DTB
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE, THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR LEAD AND ALSO THE PRODUCTION LOT HAVE BEEN RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST OF THIS LEAD AND THE PRODUCTION LOT. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFO THAT THIS LEAD WAS REMOVED DUE TO INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4137 PACER LEADS DTB BIOTRONIK GMBH AND CO. 358764

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization