CRYOCYTE FREEZING CONT 250 MLW/LAVEL POCKET
Report
- Report Number
- 1423500-2008-00481
- Event Type
- Malfunction
- Date Received
- June 11, 2008
- Date of Event
- August 17, 2006
- Report Date
- May 22, 2008
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KSE
- PMA / PMN Number
- BK950049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE IS NOT AVAILABLE FOR EVALUATION. HOWEVER, A MANUFACTURING BATCH RECORD REVIEW WILL BE PERFORMED, AND THE RESULTS WILL BE PROVIDED IN A FOLLOW-UP REPORT WHEN IT IS AVAILABLE.
THE CUSTOMER CONTACTED BAXTER HEALTHCARE TO REPORT THAT THE CRYOCYTE CONTAINER PRESENTED A CRACK AFTER IT WAS REMOVED FROM NITROGEN LIQUID. FURTHER INFORMATION RECEIVED REVEALED THAT THE TYPE OF CELLS THAT WERE STORED WERE BLOOD STEM CELLS AND BONE MARROW. THE FILL VOLUME WAS 70ML. THE DATE THE BAG FILLED WAS IN 2006. THE DATE THE BAG THAWED WAS TWO MONTHS LATER. THE DATE THE BAG RUPTURE DISCOVERED WAS THE SAME DAY. THE TYPE OF APPLICATION WERE THE CELLS INTENDED FOR WAS PATIENT USE. THE BAG BROKE DURING THAWING. THE BAG BURST NEAR THE MAIN SEALING. A FREEZING PRESS WAS USED. THE BAGS ARE PUT IN FREEZER AT A TEMPERATURE OF -80 DEGREES CELSIUS. THE BAGS ARE STORED IN METAL CASSETTES. AN OVERWRAP WAS SUED. THE BAG WAS STORED IN LIQUID NITROGEN LIQUID PHASE, AND THE BAG WAS PLACED IN THE VAPOR PHASE PRIOR TO REMOVAL FROM THE LIQUID NITROGEN STORAGE TANK FOR A MINIMUM OF 4 HRS. THE CUSTOMER IS NOT AWARE OF ANY DAMAGE OR DEVIATION FROM STANDARD PROCEDURE THAT MAY HAVE OCCURRED TO THE BAG DURING FILLING, FREEZING, STORAGE, TRANSPORT, OR THAWING. THERE HAVE BEEN NO RECENT CHANGES IN PROTOCOLS, CRITICAL EQUIPMENT/SUPPLIES OR PERSONNEL THAT MAY HAVE CONTRIBUTED TO THE BREAK. THE PROCEDURE AND THE PERSONNEL HAS BEEN THE SAME FOR 10 YEARS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CRYOCYTE FREEZING CONT 250 MLW/LAVEL POCKET | 81KSE | KSE | BAXTER HEALTHCARE | H05C08036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |