INNOVA 2100-IQ

MDR report key: 10625534 · Received October 5, 2020

Report

Report Number
9611343-2020-00004
Event Type
Death
Date Received
October 5, 2020
Date of Event
September 9, 2020
Report Date
December 14, 2020
Manufacturer
GE MEDICAL SYSTEMS SCS
Product Code
OWB
PMA / PMN Number
K052412
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6)2020 , CUSTOMER REPORTED THAT THE SYSTEM LOST MOTION WITH A PATIENT ON TABLE ON (B)(6)2020 . TO ATTEMPT TO RECOVER THE GANTRY¿¿S MOTION THE OPERATOR REBOOTED THE SYSTEM, INSTEAD OF THE EXPECTED AUTO OR MANUAL REVERSE MOTION USED TO NORMALLY CLEAR A CONTACT WITH COLLISION SENSORS. AT THE SAME PERIOD THE SEVERELY ILL PATIENT FELL INTO CARDIAC ARREST. DESPITE LIFESAVING CARDIOPULMONARY EFFORTS DURING THE REBOOT CYCLE, THE PATIENT EXPIRED. SYSTEM LOG ANALYSIS SHOWS THAT THERE WERE CASCADING USE ERRORS. 1) CUSTOMER DID NOT REBOOT THE SYSTEM FOR 2 MONTHS DESPITE WARNING MESSAGES IN EACH USE. 2) IN THIS SITUATION, THE COLLISION-SENSORS IDENTIFIED CONTACT BETWEEN TUBE AND TABLE AND THE USER DID NOT APPROPRIATELY CLEAR THEM. 3) THE POSITIONER NEEDED TO BE BACKED AWAY. 4) INSTEAD, THE OPERATOR CHOSE TO DO A COMPLETE REBOOT WHICH CANNOT CLEAR A COLLISION. THE PATIENT EXPIRED DURING THE FIRST REBOOT CYCLE AS A CONSEQUENCE OF THEIR UNDERLYING CARDIAC ARREST FOR WHICH THE ATTEMPTED LIFE SAVING MEASURES DID NOT REQUIRE FLUOROSCOPY. SUBSEQUENTLY, THERE WAS A PC BIOS BATTERY END-OF-LIFE FAILURE WHICH PREVENTED REBOOTS. THIS FAILURE COULD HAVE BEEN DETECTED DURING THE COMMON DAILY REBOOT CHECK AS RECOMMENDED IN THE PRODUCT LABELING. NO USABILITY CONCERN HAS BEEN IDENTIFIED. THE CUSTOMER WAS ABLE TO RECOVER AFTER SYSTEM RESET AND GANTRY MOTION TO CLEAR THE COLLISION. ON (B)(6)2020 , FE REPLACED THE BATTERIES AS PREVENTIVE ACTION. HE REMINDED ALSO THE CUSTOMER TO SHUT DOWN THE SYSTEM DAILY AND SHOW HOW TO CLEAR THE COLLISION. NO FURTHER ACTIONS ARE NECESSARY.

Additional Manufacturer Narrative · 1

THE CUSTOMER IS NOT ALLOWED TO PROVIDE PATIENT INFORMATION PER HIPPA REGULATIONS. CUSTOMER CONTACT ADDRESS E-MAIL WAS NOT PROVIDED. GE'S HEALTHCARE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

A PATIENT UNSTABLE ARRIVED AT THE HOSPITAL WITH SYMPTOMS OF INFARCTION. DURING THE EXAMINATION THE SYSTEM SUDDENLY SHUT DOWN, AND DESPITE SEVERAL REBOOT ATTEMPTS, IT WAS NOT POSSIBLE TO CONTINUE THE PROCEDURE. THE MEDICAL STAFF TRIED TO STABILIZE THE PATIENT, BUT HE CODED, AND DESPITE LIFE SAVING MEASURE HE DID NOT SURVIVE. GEHC BECAME AWARE OF THIS EVENT WHEN THE CUSTOMER, WHO WAS WITHOUT A SERVICE CONTRACT, GOT IN TOUCH SEEKING TO REGULARIZE HIS CONTRACT WITH THE GEHC COMMERCIAL TEAM FOR LATER SCHEDULING OF VISITS TO DIAGNOSE THE CAUSE OF THE REPORTED PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1092910 INNOVA 2100-IQ OWB GE MEDICAL SYSTEMS SCS

Patients

Seq Age Sex Outcome Treatment
1 Death