ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2008-00827
- Event Type
- Injury
- Date Received
- June 19, 2008
- Date of Event
- January 11, 2008
- Report Date
- May 23, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL: CONCLUSIONS - AT THE CONCLUSION OF THE ORIGINAL INVESTIGATION OPENED FOR THE ISSUE OF ICL VAULTING (BOTH INADEQUATE AND EXCESSIVE), IT WAS DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR BOTH INADEQUATE AND EXCESSIVE VAULT IS DIFFICULTY IN CLINICAL SIZING. IN NO CASE HAS A RETURNED LENS EXHIBITED A LENGTH MEASUREMENT OUTSIDE THE EXPECTED RANGE ACCORDING TO THE LABELED LENGTH. PRE-OPERATIVE MEASUREMENT TECHNOLOGY AVAILABLE TO CLINICIANS IN THE PAST DID NOT PROVIDE A DIRECT MEASUREMENT OF THE SULCUS, TO WHICH THE ICL SHOULD BE SIZED. THE CURRENT PRACTICE IN SELECTING AN APPROPRIATE ICL FOR A PT IS TO MEASURE WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH AND, THROUGH CORRELATIVE ALGORITHMS, PREDICT THE LENGTH OF THE ICL THAT WILL BRIDGE THE SULCUS. THIS METHOD OF SIZING BASED UPON WHITE-TO-WHITE AND ANTERIOR CHAMBER DEPTH MEASUREMENTS WAS USED IN THE FDA CLINICAL TRIAL AND IS RECOMMENDED IN THE CURRENT LABELING OF THE ICL. IF THE PREDICTED LENGTH IS TOO SHORT, THE RESULT MAY BE AN INADEQUATE VAULT. IF THE PREDICTED LENGTH IS TOO LONG, THE RESULT MAY BE AN EXCESSIVE VAULT. EVAL OF NEWLY AVAILABLE, ALTERNATE MEASURING DEVICES IS ONGOING.
IT WAS REPORTED THAT THE SURGEON INSERTED AN ICM 120V4 12.0MM IMPLANTABLE COLLAMER LENS IN 2008 AND THE LENS WAS REMOVED EIGHT DAYS LATER, DUE TO EXCESSIVE VAULT. THE PT EXPERIENCED ELEVATED IOP, NARROWING OR THE ANGLE AND SHALLOWING OF THE ACD. AN IRIDECTOMY WAS PERFORMED. THE LENS WAS EXCHANGED USING A SHORTER LENS. THIS PT HAD THE LENS EXCHANGED IN BOTH EYES. SEE MFR REPORT #2023826-2008-00828.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | ICM120V4 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | INJECTOR MODEL MSI-PF| FOAM TIP PLUNGER MODEL| CARTRIDGE MODEL UNK |