DISPOSABLE ENDOSCOPIC CLEANING BRUSH SET
Report
- Report Number
- 3001845648-2008-00002
- Event Type
- Injury
- Date Received
- June 18, 2008
- Date of Event
- May 22, 2008
- Report Date
- May 23, 2008
- Manufacturer
- COOK IRELAND LTD
- Product Code
- MNL
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
Narratives
EVAL: AS THE CLEANING BRUSH INVOLVED IN THE REPORT WAS NOT RETURNED FOR EVAL; THE REPORTED OCCURRENCE COULD NOT BE CONFIRMED. A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE CARRIED OUT AS THE LOT NUMBER OF THE CLEANING BRUSH INVOLVED IN THE REPORT IS UNK. AFTER A REVIEW OF THE ACCOUNT ORDER HISTORY, THE LOT NUMBER OF THE CLEANING BRUSH COULD NOT BE DETERMINED. A REVIEW OF THE 2-YEAR COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. THIS TYPE OF REPORT REPRESENTS AN UNUSUAL OCCURRENCE. CONCLUSION: THE FULL INVESTIGATION COULD NOT BE PERFORMED BECAUSE THE CLEANING BRUSH WAS NOT RETURNED FOR EVAL. THE CAUSE OF THE REPORT COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER WE CAN PROVIDE THE FOLLOWING COMMENTS: THE DISPOSABLE ENDOSCOPIC CLEANING BRUSH IS USED FOR CLEANING THE ACCESSORY CHANNELS OF ENDOSCOPES BETWEEN USES AND THE CLEANING BRUSH IS NOT MEDICAL DEVICE USED ON A PT. THE CLEANING BRUSH IS SUPPLIED NON-STERILE AND IS DISPOSABLE - INTENDED FOR SINGLE USE ONLY. THE MAKE AND MODEL OF THE ENDOSCOPE THAT THE DCB-DV-50 CLEANING BRUSH WAS BEING USED TO CLEAN IS UNK AT THIS TIME, THEREFORE WE CANNOT DETERMINE IF THE CLEANING BRUSH IS COMPATIBLE WITH THE ENDOSCOPE. IF THE CLEANING BRUSH WAS USED TO CLEAN AN ENDOSCOPE THAT IS NOT COMPATIBLE WITH THE CLEANING BRUSH, THIS COULD HAVE CREATED RESISTANCE DURING THE CLEANING PROCESS. IF EXCESSIVE PRESSURE WAS APPLIED TO THE CLEANING BRUSH, PERHAPS IN RESPONSE TO RESISTANCE, THIS COULD HAVE CONTRIBUTED TO BRUSH DETACHMENT INSIDE THE ACCESSORY CHANNEL OF THE ENDOSCOPE. PRIOR TO DISTRIBUTION, ALL DISPOSABLE ENDOSCOPIC CLEANING BRUSHES ARE SUBJECTED TO VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: NO CORRECTIVE ACTION IS WARRANTED AT THIS TIME BECAUSE THIS REPORT WAS UNABLE TO BE VERIFIED. THIS OBSERVATION REPRESENTS AN UNUSUAL OCCURRENCE. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
THE BRUSH HEAD OF THE DISPOSABLE CLEANING BRUSH BROKEN OFF FROM THE CATHETER INSIDE AN ENDOSCOPE. THIS WAS NOT NOTICED BY THE HOSPITAL STAFF CLEANING THE ENDOSCOPE AND THE BRUSH HEAD WAS LEFT IN THE ENDOSCOPE. AFTER THIS THE ENDOSCOPE WENT THROUGH FURTHER CLEANING PROCEDURES, THE ENDOSCOPE WAS THEN USED FOR A PROCEDURE ON A PT. THIS RESULTED IN THE BRUSH HEAD BEING PUSHED OUT OF THE ENDOSCOPE AND INTO THE PT. THE BRUSH HEAD WAS RETRIEVED FROM THE PT. THERE WERE NO ADVERSE EFFECTS ON THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DISPOSABLE ENDOSCOPIC CLEANING BRUSH SET | MNL, CLEANING BRUSHES FOR ENDOSCOPES | MNL | COOK IRELAND LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |