FDA Adverse Event Other Summary report: N

UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM

MDR report key: 1062400 · Received June 17, 2008

Report

Report Number
2122870-2008-00187
Event Type
Other
Date Received
June 17, 2008
Date of Event
May 25, 2008
Report Date
June 17, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
(K)023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE SPECIMEN WAS COLLECTED IN A BD, 13X100MM PLASTIC LITHIUM HEPARIN GEL TUBE AND WAS CENTRIFUGED ON THE AUTOMATION LINE FOR 5 MINUTES. PER CUSTOMER, THE TUBE WAS A FULL DRAW AND NO VISUAL FIBRIN OR HEMOLYSIS WAS OBSERVED. THE SAMPLE WAS NOT RE-CENTRIFUGED PRIOR TO REPEAT ANALYSIS. QC WAS IN RANGE PRIOR TO AND FOLLOWING THE EVENT. BASED ON AVAILABLE INFO THE FSE PERFORMED A PREVENTIVE MAINTENANCE (PM) AND VERIFIED THE INSTRUMENT ON 05/22/08. THE FSE WAS DISPATCHED TO THE CUSTOMER'S LAB ON 05/27/08: THE FSE FOUND COUPLER TO REAGENT GRIPPER DRIVE GEAR BROKEN AND REPLACED. THE FSE RAN A DIAGNOSTIC TESTING AND QC, AND ALL RESULTS PASSED. THE FSE VERIFIED REPAIR PER ESTABLISHED PROCEDURES AND RESULTS MEET PUBLISHED PERFORMANCE SPECIFICATIONS. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING AN ERRONEOUSLY ELEVATED ACCUTNI RESULT GENERATED BY THE UNICEL. DXI 800 ACCESS IMMUNOASSAY SYSTEM FOR A SINGLE PT. AN INITIAL ACCU TNI RESULT OF 0.55 NG/ML WAS REPORTED OUT OF THE LAB. A FRESH SAMPLE COLLECTED FROM THIS PT GAVE A RESULT OF 0.00NG/ML INITIALLY AND UPON REPEAT. THE FIRST SAMPLE WAS THEN RETESTED AND REPEATED RESULT WAS 0.00 NG/ML. THERE HAS BEEN NO REPORT OF PT INJURY, DEATH, OR CHANGE TO TREATMENT RECEIVED BY BCI TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA