FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN

MDR report key: 10623874 · Received October 2, 2020

Report

Report Number
1920898-2020-01344
Event Type
Malfunction
Date Received
October 2, 2020
Date of Event
September 7, 2020
Report Date
October 6, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED (1) 3/10CC, 12.7MM, 29G SYRINGE IN AN OPEN POLY BAG FROM LOT # 0020552. CUSTOMER STATES THAT THE NEEDLE WITH THE SHIELD WAS SEPARATED FROM THE HUB. THE RETURNED SYRINGE WAS TESTED AND THE HUB ASSEMBLY DID NOT SEPARATE FROM THE BARREL WHEN THE SHIELD WAS REMOVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0020552 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200871354] NOTED FOR CRACKED HUBS. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SYRINGE 0.3ML 29GA 1/2IN HAD THE NEEDLE SEPARATE WITH THE CAP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE HUB."

Additional Manufacturer Narrative · 1

(B)(4). DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER PHONE#: (B)(6) PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS NOT SOLD IN THE US. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYRINGE 0.3ML 29GA 1/2IN HAD THE NEEDLE SEPARATE WITH THE CAP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE HUB."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1083029 SYRINGE 0.3ML 29GA 1/2IN SYRINGE FMF BD MEDICAL - DIABETES CARE 0020552

Patients

Seq Age Sex Outcome Treatment
1 Other