FDA Adverse Event Malfunction Summary report: N

INFINITY DOCKING STATION

MDR report key: 1062302 · Received June 17, 2008

Report

Report Number
1220063-2008-00018
Event Type
Malfunction
Date Received
June 17, 2008
Date of Event
December 20, 2007
Report Date
June 16, 2008
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
MHX
Removal / Correction Number
2517967-11/16/06-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR EVAL CONFIRMED THAT THE TOP COVER OF THE DEVICES ACTUALLY SEPARATED FROM THE BASE AS REPORTED. WE ALSO FOUND THT ONE OF THE UNITS WAS MODIFIED (TOP COVER GLUED TO BASE) BY THE CUSTOMER. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICE IDENTIFIED IN THIS INCIDENT FALLS IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION.

Description of Event or Problem · 1

DRAEGER MEDICAL SYSTEMS, INC. HAS RECEIVED INFO FROM A CUSTOMER WHO HAS REPORTED THAT TWO OF OUR INFINITY DOCKING STATIONS (IDS) UNITS WERE FOUND TO HAVE SEPARATED (TOP COVER FROM THE BASE COMPONENT). NO PT INJURY WAS REPORTED. NOTE: THIS IS THE SAME REPORT AS 1220063-2008-00003 (COPY ATTACHED) EXCEPT FOR THE SECOND DEVICE WITH THAT CUSTOMER REPORTED IN THE SAME COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITY DOCKING STATION PATIENT MONITOR DOCKING STATION MHX DRAEGER MEDICAL SYSTEMS, INC. 5206110 NA

Patients

Seq Age Sex Outcome Treatment
1 NI