FDA Adverse Event Malfunction Summary report: N

CUSTOM COMBI SET

MDR report key: 1062300 · Received June 18, 2008

Report

Report Number
8030665-2008-00033
Event Type
Malfunction
Date Received
June 18, 2008
Date of Event
May 19, 2008
Report Date
June 18, 2008
Manufacturer
REYNOSA MANUFACTURING
Product Code
FJK
PMA / PMN Number
K962081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE HISTORY LOT REVIEW: NO INDICATION OF THE REPORTED DEFECT WAS FOUND DURING THE DHR REVIEW. LOT MEETS ALL RELEASED CRITERIA. SAMPLE ANALYSIS: WE RECEIVED A ARTERIAL LINE WITHOUT THE ORIGINAL PACKAGING. DURING VISUAL ANALYSIS, WE COULD DETECT A SEPARATION AT SALINE TUBING TO "T" CONNECTOR. NO SOLVENT ON TUBING. CONCLUSION: THE COMPLAINT IS CONFIRMED, CORRECTIVE ACTION: THE ROOT CAUSE COULD BE ATTRIBUTED TO INSUFFICIENT AMOUNT OF SOLVENT USED DURING THE BONDING OPERATION. FMC REYNOSA HAS DESIGNATED RESOURCES, STANDARDIZING THE BUSHING FROM 1/4 TO 3/8 INCH. DESCRIPTION OF THE CHANGE FOR THIS NEW CONFIGURATION IS DOCUMENTED. FMC REYNOSA WILL CONTINUE TO MONITOR OCCURRENCES, AND TO IMPLEMENT CHANGES IN ORDER TO ELIMINATE THIS FAILURE MODE.

Description of Event or Problem · 1

AN ELECTRONIC REPORT WAS RECEIVED FROM A HEMODIALYSIS USER FACILITY. THE FACILITY HAS REPORTED THAT THE SALINE LINE SEPARATED FROM THE BLOOD TUBING AFTER THE TREATMENT HAD STARTED. THE FACILITY HAS BEEN CONTACTED WHERE IT HAS BEEN LEARNED THAT AT THE START OF DIALYSIS, THE SALINE LINE SEPARATED FROM THE ARTERIAL LINE. THE LINE WAS CLAMPED RIGHT AWAY. THE RN REPORTED NO BLOOD LOSS. ONCE THE EVENT WAS DETECTED, A NEW ARTERIAL LINE WAS SET UP ON THE DIALYSIS MACHINE AND THE PRESCRIBED TREATMENT WAS ABLE TO BE COMPLETED WITHOUT FURTHER INCIDENCE. THERE IS NO REPORT OF INJURY OR ILL EFFECT. A SAMPLE HAS BEEN SAVED FOR AN EVAL. THE PT WAS DISCHARGED TO HOME WHEN THE TREATMENT WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CUSTOM COMBI SET HEMODIALYSIS BLOODLINES FJK REYNOSA MANUFACTURING NA 7LR073

Patients

Seq Age Sex Outcome Treatment
1 NA