Malfunction
Summary report: N
INFINITY DOCKING STATION
MDR report key: 1062297
·
Received June 18, 2008
Report
- Report Number
- 1220063-2008-00019
- Event Type
- Malfunction
- Date Received
- June 18, 2008
- Date of Event
- May 15, 2008
- Report Date
- June 18, 2008
- Manufacturer
- DRAEGER MEDICAL SYSTEMS, INC. (IT/M)
- Product Code
- MHX
- Removal / Correction Number
- 2517967-11/16/06-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
OUR EVAL CONFIRMED THAT THE TOP COVER OF THE DEVICE ACTUALLY SEPARATED FROM THE BASE AS REPORTED. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICE IDENTIFIED IN THIS INCIDENT FALLS IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION NUMBER: (B) (4).
Description of Event or Problem · 1
WHEN TAKING THE PT MONITOR INFINITY DELTA FROM THE INFINITY DOCKING STATION (IDS) IT FELL DOWN. THE TOP COVER OF THE IDS SEPARATED FROM THE BOTTOM (THE NUTS WERE TORN OFF). NO PERSONS WERE INJURED. (B) (4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITY DOCKING STATION | PATIENT MONITOR DOCKING STATION | MHX | DRAEGER MEDICAL SYSTEMS, INC. (IT/M) | 5206110 | 5466271863 |