FDA Adverse Event Malfunction Summary report: N

INFINITY DOCKING STATION

MDR report key: 1062297 · Received June 18, 2008

Report

Report Number
1220063-2008-00019
Event Type
Malfunction
Date Received
June 18, 2008
Date of Event
May 15, 2008
Report Date
June 18, 2008
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC. (IT/M)
Product Code
MHX
Removal / Correction Number
2517967-11/16/06-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR EVAL CONFIRMED THAT THE TOP COVER OF THE DEVICE ACTUALLY SEPARATED FROM THE BASE AS REPORTED. THIS IS A KNOWN SUPPLIER ISSUE AND WE HAVE IMPLEMENTED SUPPLIER CORRECTIVE ACTION AND INITIATED A FIELD CORRECTION. THE DEVICE IDENTIFIED IN THIS INCIDENT FALLS IN THE RANGE OF THE POTENTIALLY DEFECTIVE DEVICES OF THE FIELD CORRECTION NUMBER: (B) (4).

Description of Event or Problem · 1

WHEN TAKING THE PT MONITOR INFINITY DELTA FROM THE INFINITY DOCKING STATION (IDS) IT FELL DOWN. THE TOP COVER OF THE IDS SEPARATED FROM THE BOTTOM (THE NUTS WERE TORN OFF). NO PERSONS WERE INJURED. (B) (4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITY DOCKING STATION PATIENT MONITOR DOCKING STATION MHX DRAEGER MEDICAL SYSTEMS, INC. (IT/M) 5206110 5466271863

Patients

Seq Age Sex Outcome Treatment
1