FDA Adverse Event
Malfunction
Summary report: N
PREICSION LINK
MDR report key: 1062263
·
Received June 17, 2008
Report
- Report Number
- 2954323-2008-02120
- Event Type
- Malfunction
- Date Received
- June 17, 2008
- Date of Event
- May 23, 2008
- Report Date
- June 17, 2008
- Manufacturer
- ABBOTT DIABETES CARE INC, USA
- Product Code
- NBW
- Removal / Correction Number
- Z-0709-2007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE IS A KNOWN MALFUNCTION WITH THE PRECISION LINK SOFTWARE THAT CAN LEAD TO INCORRECT TRENDING OF RESULTS. THIS OCCURS WHEN RESULTS OBTAINED ON A METER WITH INCORRECT DATE AND TIME ARE UPLOADED TO A COMPUTER WITH PRECISION LINK SOFTWARE. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH ADC FA21DEC2006 LETTER.
Description of Event or Problem · 1
A CUSTOMER REPORTED RECEIVING AN ERROR 6 DISPLAYING ON THEIR PRECISION XTRA BLOOD GLUCOSE METER. IT WAS THEN ADDITIONALLY IDENTIFIED BY ADC CUSTOMER SERVICE THAT THE DATE AND TIME SETTINGS IN THEIR METER WERE NOT PROPERLY SET, AND THEY REPORTED TO BE A USER OF THE PRECISION LINK DATA MANAGEMENT SYSTEM. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PREICSION LINK | DATA MANAGEMENT SYSTEM | NBW | ABBOTT DIABETES CARE INC, USA | NA | 41938 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |