KENTROX RV-S 65
Report
- Report Number
- 1028232-2008-00655
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Report Date
- May 21, 2008
- Manufacturer
- BIOTRONIK GMBH AND CO.
- Product Code
- LWS
- PMA / PMN Number
- P980023
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE ANALYSIS OF THE LEADS TESTED THE MECHANICAL AND ELECTRICAL PROPERTIES. THE SELOX SR SHOWED A DEFORMATION (SQUASHING) OF THE OUTER HELIX AND A CONDUCTOR FRACTURE OF THE INNER CONDUCTOR HELIX ABOUT 2 CM DISTAL OF THE LIGATURE. THIS DAMAGE MUST BE REGARDED THE CAUSE OF THE SUDDEN INCREASE IN THE PACING IMPEDANCE REPORTED ABOUT IN THE COMPLAINT. FURTHERMORE, DAMAGE TO THE INNER INSULATION JUST DISTAL FROM THIS SPOT COULD BE DETECTED. THIS DAMAGE MANIFESTATION IS, WITH HIGH PROBABILITY, DUE TO THE OCCURRENCE OF STRONG LOCAL PRESSURE LOADS AT THE LEAD BODY. THE CLOSE VICINITY OF THE FRACTURE OF THE INNER CONDUCTOR HELIX TO THE INTRODUCER AID (EFH) AND THE CHARACTERISTICS OF THE DAMAGE MANIFESTATION LEAD TO THE ASSUMPTION THAT THE LEAD WAS CLAMPED BETWEEN THE EFH AND THE BONES. JAMMING OF THE LEAD BETWEEN THE CLAVICLE AND THE FIRST RIB (SUBCLAVIAN CRUSH SYNDROME) SHOULD HERE BE CONSIDERED. DUE TO THE DYNAMIC STRESS ON THE CLAMPED LEAD EXERTED WHEN THE PT IS MOVING, ABRASION TRACES FORMED, AND DUE TO EXCESSIVE REVERSED BENDING, PRESSURE AND PULLING STRESSES, THE INTERNAL HELIX BROKE. DIAGNOSTIC IMAGES OF THE MENTIONED BODY REGION WERE NOT AVAILABLE FOR A FURTHER INVESTIGATION OF THIS SITUATION. THE ABRASION TRACES ON THE LEAD BODIES OF BOTH LEADS ARE PROBABLY DUE TO CONTINUOUS LOCAL FRICTION OF THE LEADS AGAINST EACH OTHER. THE CUTS IN THE OUTER INSULATION AND THE BLOOD TRACES WITHIN THE LEAD KENTROX RV-S ARE PROBABLY DUE TO THE EXPLANTATION PROCEDURE. THE ANALYSIS WAS UNABLE TO FIND ANY MFG ERROR OR MATERIAL DEFECT.
OUS MDR. A SUDDEN INCREASE IN THE LEAD IMPEDANCE TO ABOUT 3 KOHM WITH SUSPICION OF LEAD FRACTURE WAS REPORTED AFTER AN IMPLANTATION TIME OF ABOUT 6 WEEKS. THE LEADS AND THE ICD WERE EXPLANTED. ALSO REMOVED: LEXOS VR-T, MDR 1028232-2008-00656. SELOX SR 60, MDR 1028232-2008-00657.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KENTROX RV-S 65 | ICD LEAD | LWS | BIOTRONIK GMBH AND CO. | 338148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |