SELOX SR 60
Report
- Report Number
- 1028232-2008-00657
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Report Date
- May 21, 2008
- Manufacturer
- BIOTRONIK GMBH AND CO.
- Product Code
- DTB
- PMA / PMN Number
- P950037
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THE ANALYSIS OF THE LEADS TESTED THE MECHANICAL AND ELECTRICAL PROPERTIES. THE SELOX SR SHOWED A DEFORMATION (SQUASHING) OF THE OUTER HELIX AND A CONDUCTOR FRACTURE OF THE INNER CONDUCTOR HELIX ABOUT 2 CM DISTAL OF THE LIGATURE. THIS DAMAGE MUST BE REGARDED THE CAUSE OF THE SUDDEN INCREASE IN THE PACING IMPEDANCE REPORTED ABOUT IN THE COMPLAINT. FURTHERMORE, DAMAGE TO THE INNER INSULATION JUST DISTAL FROM THIS SPOT COULD BE SEEN. THIS DAMAGE MANIFESTATION IS, WITH HIGH PROBABILITY, DUE TO THE OCCURRENCE OF STRONG LOCAL PRESSURE LOADS ON THE LEAD BODY. THE CLOSE VICINITY OF THE FRACTURE OF THE INNER CONDUCTOR HELIX TO THE INTRODUCER AID (EFH) AND THE CHARACTERISTICS OF THE DAMAGE MANIFESTATION LEAD TO THE ASSUMPTION THAT THE LEAD WAS CLAMPED BETWEEN THE EFH AND THE BONES. JAMMING OF THE LEAD BETWEEN THE CLAVICLE AND THE FIRST RIB (SUBCLAVIAN CRUSH SYNDROME) SHOULD BE CONSIDERED. DUE TO THE DYNAMIC STRESS ON THE CLAMPED LEAD EXERTED WHEN THE PT IS MOVING, ABRASION TRACES FORMED, AND DUE TO EXCESSIVE REVERSED BENDING, PRESSURE AND PULLING STRESSES, THE INTERNAL HELIX BROKE.
OUS MDR. A SUDDEN INCREASE IN THE LEAD IMPEDANCE TO ABOUT 3 KOHM WITH SUSPICION OF LEAD FRACTURE WAS REPORTED AFTER AN IMPLANTATION TIME OF ABOUT 6 WEEKS. THE LEADS AND THE ICD WERE EXPLANTED. ALSO REMOVED: LEXOS VR-T, MDR 1028232-2008-00656. KENTROX RV-S 65, MDR 1028232-2008-00655.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SELOX SR 60 | PACER LEAD | DTB | BIOTRONIK GMBH AND CO. | 343082 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |