FDA Adverse Event Malfunction Summary report: N

LOW PROFILE U-JOINT DRIVER

MDR report key: 1062157 · Received June 13, 2008

Report

Report Number
1719045-2008-00077
Event Type
Malfunction
Date Received
June 13, 2008
Date of Event
April 22, 2008
Report Date
April 22, 2008
Manufacturer
SYNTHES (USA)
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SUBJECT DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED. SUBJECT DEVICE WAS REC'D AND EVALUATED. THE BROKEN SURFACE DOES NOT SHOW ANY ANOMALIES. MECHANICAL OVERLOADING POSSIBLY LEAD TO THE BREAKAGE. REVIEW OF THE MFG RECORDS FOUND NO COMPLAINT-RELATED ISSUES.

Description of Event or Problem · 1

THE STARDRIVE ON A LOW PROFILE U-JOINT DRIVER BROKE OFF INSIDE THE HEAD OF A SCREW. THE BROKEN FRAGMENT COULD NOT BE REMOVED, SO THE SURGEON PUT BONE WAX OVER THE BROKEN FRAGMENT. THE SURGEON HAS NO PLANS TO EXPLANT THE BROKEN FRAGMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LOW PROFILE U-JOINT DRIVER LOW PROFILE U-JOINT DRIVER LXH SYNTHES (USA) NA 2331810

Patients

Seq Age Sex Outcome Treatment
1 NI SCREW| BONE WAX