FDA Adverse Event
Malfunction
Summary report: N
LOW PROFILE U-JOINT DRIVER
MDR report key: 1062157
·
Received June 13, 2008
Report
- Report Number
- 1719045-2008-00077
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Date of Event
- April 22, 2008
- Report Date
- April 22, 2008
- Manufacturer
- SYNTHES (USA)
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
SUBJECT DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED. SUBJECT DEVICE WAS REC'D AND EVALUATED. THE BROKEN SURFACE DOES NOT SHOW ANY ANOMALIES. MECHANICAL OVERLOADING POSSIBLY LEAD TO THE BREAKAGE. REVIEW OF THE MFG RECORDS FOUND NO COMPLAINT-RELATED ISSUES.
Description of Event or Problem · 1
THE STARDRIVE ON A LOW PROFILE U-JOINT DRIVER BROKE OFF INSIDE THE HEAD OF A SCREW. THE BROKEN FRAGMENT COULD NOT BE REMOVED, SO THE SURGEON PUT BONE WAX OVER THE BROKEN FRAGMENT. THE SURGEON HAS NO PLANS TO EXPLANT THE BROKEN FRAGMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LOW PROFILE U-JOINT DRIVER | LOW PROFILE U-JOINT DRIVER | LXH | SYNTHES (USA) | NA | 2331810 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | SCREW| BONE WAX |