FDA Adverse Event Malfunction Summary report: N

MODEL 91341, TELEMETRY TRANSMITTER 1400 MHZ

MDR report key: 1062156 · Received June 13, 2008

Report

Report Number
3023361-2008-00007
Event Type
Malfunction
Date Received
June 13, 2008
Date of Event
March 9, 2007
Report Date
June 13, 2008
Manufacturer
SPACELABS HEALTHCARE
Product Code
DSI
PMA / PMN Number
K925510
Removal / Correction Number
Z-1202-6
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED AN INCIDENT IN WHICH, UPON BEING POWERED UP, A TELEMETRY TRANSMITTER WOULD BEGIN TRANSMITTING ON THE WRONG RADIO FREQUENCY CHANNEL (I.E. A FREQUENCY OTHER THAN THE FREQUENCY ON WHICH IT HAD BEEN SET TO TRANSMIT) WHEN THIS OCCURS, IF ANOTHER TELEMETRY RECEIVER AT THE SITE HAS BEEN TUNED TO RECEIVE TRANSMISSIONS ON THIS OTHER (WRONG) CHANNEL, THE PATIENT'S WAVEFORM WOULD BE DISPLAYED ON THE CENTRAL MONITOR AS HAVING BEEN TRANSMITTED ON THIS WRONG CHANNEL. PRIOR TO RECEIPT OF THIS COMPLAINT, WE HAD ALREADY INITIATED A RECALL TO ADDRESS PREVIOUSLY REPORTED INCIDENTS OF CHANNEL SWITCHING. AT THE TIME OF THIS COMPLAINT, THIS CUSTOMER'S PRODUCTS HAD NOT YET BEEN UPDATED IN CONNECTION WITH THE RECALL CORRECTIVE ACTION. THIS RECALL IS STILL IN PROCESS. ADDITIONAL SERIAL NUMBER: (B)(4).

Description of Event or Problem · 1

BIOMED (B)(6) REPORTED THAT A PT WEARING A MODEL 91341 TELEMETRY TRANSMITTER THAT WAS SET TO CHANNEL #1688 WAS DISCHARGED AND THE BATTERY WAS REMOVED FROM THE TRANSMITTER. ANOTHER PT WEARING A DIFFERENT TELEMETRY TRANSMITTER SET TO CHANNEL #1689, WENT TO A PROCEDURE AND THE BATTERY WAS REMOVED FROM THE TRANSMITTER. WHEN THE PT RETURNED, THE BATTERY WAS REINSERTED INTO THE TRANSMITTER WHICH THEN IMPROPERLY BEGAN TRANSMITTING ON CHANNEL #1688. (B)(6) INDICATED THE BATTERY WAS REMOVED AND THE TRANSMITTER WAS TESTED, THE SIGNAL REMAINED ON THE IMPROPER CHANNEL, #1688. MR. MASON THEN RETESTED THE TELEMETRY TRANSMITTER ABOUT TWO HOURS LATER AND IT TRANSMITTED ON THE PROPER CHANNEL, #1689.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MODEL 91341, TELEMETRY TRANSMITTER 1400 MHZ NONE DSI SPACELABS HEALTHCARE 91341

Patients

Seq Age Sex Outcome Treatment
1