FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 1062146 · Received June 13, 2008

Report

Report Number
1220908-2008-01283
Event Type
Malfunction
Date Received
June 13, 2008
Report Date
May 29, 2008
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED MALFUNCTION WAS OBSERVED AND ATTRIBUTED TO AN OPEN WIRE WITHIN THE MULTIFUNCTION CABLE. THE MULTIFUNCTION CABLE WAS REPLACED TO REMEDY THE PROBLEM CONDITION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF FAILURES OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WAS UNABLE TO DISCHARGE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA