FDA Adverse Event Injury Summary report: N

DROP ARM COMMODE CHAIR MODEL

MDR report key: 10619377 · Received October 1, 2020

Report

Report Number
MW5097021
Event Type
Injury
Date Received
October 1, 2020
Date of Event
July 10, 2020
Report Date
September 29, 2020
Manufacturer
MEDLINE INDUSTRIES
Product Code
INB
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

BROKE LEG AND HIP; INTRODUCTION: MY NAME IS (B)(6). I LIVE IN (B)(6). I AM RETIRED AND HAVE BEEN SINCE BECOMING PARAPLEGIC 22 YEARS AGO. I HAVE SUCCESSFULLY TRANSFERRED ONTO AND OFF OF A COMMODE CHAIR LIKE THE ONE IN THE PHOTO FOR MOST OF THIS TIME. A COMMODE CHAIR IS BOTH A TOILET AND A SURFACE TO USE FOR ME TO TAKE A SHOWER. ON (B)(6) 2020 WHILE TRANSFERRING FROM THE SHOWER BENCH ONTO THE COMMODE CHAIR, THE RIGHT DROP-ARM FAILED TO SECURELY LOCK INTO PLACE AND I FELL BREAKING BOTH BONES IN MY LEFT LEG. I WEIGH ABOUT (B)(6) LBS. THE WEIGHT RATING IS 250 LBS. THE UNIT IS MADE BY (B)(4), MODEL NUMBER MDS89668 AND IS FROM LOT NUMBER 108 9010012. EVERY TRANSFER IS COMPLETED AS A 90 DEGREE ROTATION WITH ONE HAND ON EACH SURFACE AND REQUIRE A COUPLE OF SMALL HOPS. DETAILED DESCRIPTION: THE DROP-ARM ON THE COMMODE CHAIR SHOULD BE SNUG ENOUGH IN THE REAR WHERE IT ATTACHES TO THE FRAME SO THAT WHEN THE HANDLE IS RAISED THE LATCH MEETS THE FRAME IN THE CENTER AND THE EDGES OF THE LATCH LAND ON THE PINS DESIGNED TO HOLD THE HANDLE UP. THE IMAGE ATTACHED SHOWS HOW THE DROP-ARM HAS TOO MUCH SWAY AND CAN BE BROUGHT UP SO THAT THE ARM IS OFF CENTER AND COMES TO REST ON THE OUTSIDE PIN. TO TAKE A SHOWER, THERE ARE TWO TRANSFERS. THE FIRST TRANSFER IS FROM MY WHEELCHAIR TO THE COMMODE CHAIR. THE DROP-ARM IN QUESTION IS DOWN FOR THE TRANSFER THEN BROUGHT UP AFTER I'M ON THE COMMODE CHAIR. THE DAY I FELL, WHEN I BROUGHT THE HANDLE UP INTO POSITION IT CAME TO REST IN THE OUTSIDE POSITION BUT I COULDN'T SEE THE PIN AREA BECAUSE I LAY A TOWEL OVER THE SEAT, AND IT COVERS THE PINS WHILE STILL ALLOWING FOR THE DROP-ARM TO RAISE AND LOCK INTO PLACE. I DON'T PAY MUCH ATTENTION TO THIS NORMAL OPERATION. WHEN I TRANSFER ONTO THE COMMODE CHAIR I PUT MY HANDS ON THE WHEELCHAIR AND THE COMMODE CHAIR. THE SECOND TRANSFER IS INTO THE SHOWER. FOR THIS MOVE I LOWER THE LEFT DROP-ARM. WHEN I TRANSFER ONTO THE SHOWER BENCH I PUT MY HANDS ON THE COMMODE SEAT AND THE SHOWER BENCH. AFTER MY SHOWER I COMPLETE THE FIRST OF TWO HOPS OF MY TRANSFER ONTO THE COMMODE CHAIR. THEN FOR THE OTHER HALF I REACH TO THE FAR DROP-ARM AND USE IT TO COMPLETE THE TRANSFER. AS I PUT WEIGHT ON THAT ARM IT SLIPPED OFF THE PIN AND AS MY WEIGHT WAS ON MY RIGHT ARM, WHEN THE DROP-ARM WENT DOWN SO DID I. IT'S A PHYSICS ISSUE. GRAVITY ALWAYS WINS. IF THE DROP-ARM HAD BEEN SECURELY FASTENED, IT WOULD'VE SUPPORTED MY WEIGHT AS IT ALWAYS HAS. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1075734 DROP ARM COMMODE CHAIR MODEL ATTACHMENT, COMMODE, WHEELCHAIR INB MEDLINE INDUSTRIES MDS89668 108 9010012

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| O| R