WALL MOUNT INFANT WARMER
Report
- Report Number
- 9611451-2008-00367
- Event Type
- Malfunction
- Date Received
- June 16, 2008
- Report Date
- May 22, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD.
- Product Code
- FMT
- PMA / PMN Number
- K971461
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
ADDITIONAL LOT NUMBERS: 051012001096, 051012001098, 051012001100, 051012001095, 041201001218, 051017001123, 051012001099. ADDITIONAL MANUFACTURING DATES: 10/12/05, 10/12/05, 10/12/05, 10/12/05. 12/01/04, 10/17/05, 10/12/05. AN INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICES HAVE NOT BEEN RETURNED FOR INVESTIGATIONS. RESULTS: IT WAS REPORTED THAT THE WARMER HEADS OF THE INFANT WARMER BEGAN TO BREAK DOWN DUE TO THE USE OF PRO FIL # 13 ULTRA PLUS. IT IS A CLEANING SOLUTION THAT CONTAINS CHEMICAL INGREDIENT THAT IS NOT RECOMMENDED FOR USE ON THE WARMER HEAD, AS SPECIFIED ON THE OPERATING MANUAL OF THE INFANT WARMER. THE OPERATING MANUAL STATES THAT CLEANING OF ALL PARTS OF THE INFANT WARMER AND ACCESSORIES SHOULD BE DONE AT NORMAL ROOM TEMPERATURE, AROUND 23 DEGREE C, AND TO AVOID CLEANING SOLUTIONS CONTAINING AROMATIC HYDROCARBONS, AMMONIA, AMINES OR AQUEOUS SOLUTIONS AS THESE MAY CAUSE CHEMICAL REACTIONS TO THE POLYCARBONATE PLASTIC MATERIAL. CONCLUSION: REPORTED FAULT WAS CONFIRMED AS USE ERROR. THE DEGRADATION OF THE INFANT WARMER HEAD, DUE TO THE USE OF NON-COMPATIBLE CLEANING MATERIALS, IS A KNOWN PROBLEM. WE HAVE AN OPEN CAPA TO EXAMINE FURTHER COMPATIBLE CLEANING SOLUTIONS.
A HOSPITAL IN ANOTHER COUNTRY, REPORTED THAT EIGHT IW980JEU WALL MOUNT INFANT WARMER HEADS BEGAN TO BREAK DOWN DUE TO THE USE OF DISINFECTANT SOLUTION CONTAINING ACTIVE INGREDIENT THAT IS LISTED IN THE INFANT WARMER OPERATING MANUAL AS NOT RECOMMENDED FOR USE. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WALL MOUNT INFANT WARMER | FMT | FISHER & PAYKEL HEALTHCARE LTD. | IW980JEU | 051012001096 |