FDA Adverse Event Injury Summary report: N

HERCULES 6700B

MDR report key: 1061907 · Received June 13, 2008

Report

Report Number
1061907
Event Type
Injury
Date Received
June 13, 2008
Date of Event
February 28, 2008
Report Date
June 11, 2008
Manufacturer
SKYTRON
Product Code
FQO
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THIS PT WAS ADMITTED TO HOSPITAL IN 2008 FOR A PROLONGED SURGICAL PROCEDURE THAT LASTED 12 HOURS. AT APPROX 9-10 HOURS INTO THE PROCEDURE, IT WAS NOTED THAT THE LOWER EXTREMITY PORTION OF THE TABLE HAD BEEN INADVERTENTLY ADJUSTED IN THE FULL DOWN POSITION, SO THAT THE LOWER EXTREMITIES WERE BENT DOWN ALMOST 90 DEGREES AT THE KNEE. PT DEVELOPED A SERIOUS INJURY: A FOUR-COMPARTMENT SYNDROME OF THE RIGHT LEG. THREE DAYS LATER, HE UNDERWENT A FOUR-COMPARTMENT FASCIOTOMY OF THE RIGHT LEG WITH COMPLETE RELEASE OF THE SKIN AND FASCIA OF THE INVOLVED COMPARTMENTS. PT WAS DISCHARGED HOME TWELVE DAYS LATER. THIS EVENT WAS REPORTED TO DEPARTMENT OF HEALTH TWO WEEKS LATER. TWO MONTHS LATER, THE O.R. TEAM AND RISK MGMT IDENTIFIED THE ONE-TOUCH PENDANT CONTROL OF THE SURGICAL TABLE (SKYTRON HERCULES 6700B) TO BE AN ISSUE IN REGARD TO PT SAFETY FOR THE INADVERTENT CHANGE IN THE PT POSITION THAT CAUSED A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HERCULES 6700B SURGICAL TABLE FQO SKYTRON 6700B NA

Patients

Seq Age Sex Outcome Treatment
1 18 YR Hospitalization| R