FDA Adverse Event Injury Summary report: N

WORKMATE CLARIS SYSTEM AMPLIFIER

MDR report key: 10617670 · Received October 1, 2020

Report

Report Number
2184149-2020-00165
Event Type
Injury
Date Received
October 1, 2020
Date of Event
September 28, 2020
Report Date
October 26, 2020
Manufacturer
ST. JUDE MEDICAL, INC.
Product Code
DQK
UDI-DI
05415067001696
PMA / PMN Number
K151911
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ONE WORKMATE CLARIS AMPLIFIER WAS RECEIVED FOR EVALUATION. THE RETURNED WORKMATE CLARIS AMPLIFIER WAS POWERED ON BUT NO COMMUNICATION WAS ESTABLISHED WITH THE TEST STANDARD WORKMATE CLARIS COMPUTER. THE SINGLE BOARD COMPUTER (SBC) WAS TEMPORARILY REPLACED WITH A KNOWN GOOD SBC BOARD AND SUCCESSFUL COMMUNICATION WAS ESTABLISHED WITH THE TEST STANDARD WORKMATE CLARIS COMPUTER. USING THE TEST STANDARD SBC BOARD, THE AMPLIFIER WAS RUN FOR FEW HOURS AND NO INTERRUPTED OR LOSS OF COMMUNICATION WAS OBSERVED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Description of Event or Problem · 1

DURING THE PROCEDURE WITH THE PATIENT ON THE TABLE, AN "AMPLIFIER DISCONNECTED" ERROR WAS NOTED AND THE CASE WAS CANCELLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1076061 WORKMATE CLARIS SYSTEM AMPLIFIER COMPUTER, DIAGNOSTIC, PROGRAMMABLE DQK ST. JUDE MEDICAL, INC. H700169 6472429 05415067001696

Patients

Seq Age Sex Outcome Treatment
1