DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00063
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Date of Event
- May 18, 2008
- Report Date
- May 18, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K944093
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
NO FURTHER EVAL OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS SLIPPING HM WASH PUMP DUE TO LACK OF ROUTINE MAINTENANCE BY THE OPERATOR. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER CORRECTED THE PROBLEM WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS INC TECHNICAL SOLUTIONS CENTER REPRESENTATIVE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. SEE SCANNED PAGES.
A FALSELY ELEVATED TROPONIN I RESULT OF 0.91/0.79 NG/ML WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON THE SAME ANALYZER AND A RESULT OF 0.02 NG/ML WAS OBTAINED. THE SAMPLE WAS REPEATED ON AN ALTERNATE ANALYZER AND A RESULT OF 0.04 NG/ML WAS OBTAINED. IT IS UNKNOWN IF PT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS SLIPPING HM WASH PUMP DUE TO LACK OF ROUTINE MAINTENANCE BY THE OPERATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | RXLMAXRH | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |