ALLEN MEDICAL SYSTEMS
Report
- Report Number
- 1221538-2008-00003
- Event Type
- Other
- Date Received
- June 13, 2008
- Date of Event
- May 1, 2008
- Report Date
- May 15, 2008
- Manufacturer
- ALLEN MEDICAL SYSTEMS, INC.
- Product Code
- FWZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
UPON RETURN OF THE DEVICE, ENGINEERING EVAL REVEALED THE PUSH BUTTON RELEASE WAS LODGED INSIDE THE SUPPORT ROD AND RENDERED INOPERABLE. THE END STOP ON THE SUPPORT ROD WAS ALSO MISSING. PER THE INSTRUCTIONS FOR USE, THE CHAIR IS TO BE CHECKED BEFORE EACH USE. IF THE TWO COMPONENTS WERE FUNCTIONAL ON THE UNIT AS THEY ARE DESIGNED AND SOLD, THE SUPPORT RODS COULD NOT HAVE COME FREE FROM THE CHAIR FRAME AND THE INCIDENT WOULD NOT HAVE OCCURRED. THE EVAL CONCLUDED THAT THE INCIDENT WAS THE RESULT OF CUSTOMER MISUSE. THIS IS DUE TO THE FACT THAT THE BEACH CHAIR WAS USED FOR THIS PROCEDURE AFTER THE DEVICE WAS ALREADY DAMAGED.
IN 2008, REP ALLEN WAS CONTACTED BY A DISTRIBUTOR ABOUT A PATIENT-INVOLVED INCIDENT INVOLVING ONE OF OUR BEACH CHAIRS. ALLEN MEDICAL RECEIVED A COPY OF THE FACILITY'S VOLUNTARY MEDWATCH FILING FROM THE DISTRIBUTOR. ACCORDING TO THE REPORT, A PT - WHO WAS UNDER GENERAL ANESTHESIA - WAS BEING POSITIONED IN A BEACH CHAIR PRIOR TO THE START OF A PROCEDURE. WHEN THE USER ATTEMPTED TO EXTEND THE HEADREST SUPPORT RODS, THE RODS EXTENDED BEYOND THE LOCKING POSITION AND CAME FREE FROM THE CHAIR FRAME. THIS LEFT THE PT'S HEAD UNSUPPORTED. THE PT'S HEAD WAS KEPT SUPPORTED BY STAFF MEMBERS ON HAND AND THERE WAS NO INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLEN MEDICAL SYSTEMS | BEACH CHAIR/DIRECT SHOULDER POSITIONER | FWZ | ALLEN MEDICAL SYSTEMS, INC. | F-SCBER-A3 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |