FDA Adverse Event Other Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 1061711 · Received June 13, 2008

Report

Report Number
2955842-2008-01073
Event Type
Other
Date Received
June 13, 2008
Date of Event
May 19, 2008
Report Date
June 13, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING FOUND NO ERRORS IN THE DA VINCI S SYSTEM LOG. PERFORMANCE TESTS FOR THE SPECIFIC PT SIDE MANIPULATOR (PSM) ARM WERE CONDUCTED AND THE CUSTOMER REPORTED PROBLEM COULD NOT BE DUPLICATED. THE PT SIDE MANIPULATOR IS AN INSTRUMENT ARM LOCATED ON THE PT SIDE CART, THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. ENGINEERING CONCLUDED THAT THE INSTRUMENT INSERTED IN THE PSM WAS MOST LIKELY AT THE END OF THE CANNULA ACCESSORY AND THE "JERKY" MOTION WAS DUE TO THIS POOR POSITIONING. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S BEATING HEART TECAB PROCEDURE, APPROX FIVE HOURS INTO THE PROCEDURE, THE SURGEON STATED THAT A PT SIDE MANIPULATOR (PSM) ARM FELT "JERKY". THE SITE CONVERTED TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO FINISH THE PLANNED SURGERY. NO PT HARM WAS REPORTED. THE CUSTOMER CONTACTED ISI TECHNICAL SUPPORT AFTER THE PROCEDURE WAS COMPLETE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS2000 A5.1P4

Patients

Seq Age Sex Outcome Treatment
1 Other INSTRUMENTS| DA VINCI S SURGICAL SYSTEM| ACCESSORIES