- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING CONCLUDED THAT THE SYSTEM ERROR CODE #297 EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH A CONFIGURED EMBEDDED SERIALIZER SETUP JOINT (CFG, ESSJ). THE EMBEDDED SERIALIZER FOR SETUP-JOINT IS THE PRINTED CIRCUIT ASSEMBLY (PCA) INSIDE A SYSTEM ARM THAT MONITORS THE POTENTIOMETER FOR EACH OF THE FOUR JOINTS AND THEIR ASSOCIATED BACKUP POTENTIOMETERS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED CFG ESSJ AND RECONNECTING THE LOW VOLTAGE DIFFERENTIAL SIGNAL (LVDS) CABLE TP THE PCA. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PT. SYSTEM ERROR CODE #23 INDICATES A COMMUNICATION FAILURE BETWEEN THE SURGEON CONSOLE AND THE PT SIDE CART AND THIS COMMUNICATION FAILURE CAUSED AN INCOMPLETE PCA SOFTWARE DOWNLOAD WHICH CREATED THE PERSISTENT FAULT CONDITION ERROR CODE OF #297. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED WHILE SETTING UP THE DA VINCI S SURGICAL SYSTEM FOR A PROSTATECTOMY PROCEDURE THE CUSTOMER EXPERIENCED SYSTEM ERROR CODE #297. THE CUSTOMER RESTARTED THE SYSTEM, HOWEVER, SYSTEM ERROR CODE #297 RECURRED UPON POWER UP. THE PT WAS UNDER ANESTHESIA AND THE PORT INCISIONS HAD BEEN CREATED WHEN THE SITE DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO FINISH THE PLANNED SURGERY. NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI S SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL, INC. | IS2000 A5. 1P4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | ACCESSORIES| DA VINCI S SURGICAL SYSTEM| INSTRUMENTS |