FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1061654 · Received June 17, 2008

Report

Report Number
3003464075-2008-00262
Event Type
Injury
Date Received
June 17, 2008
Date of Event
May 21, 2008
Report Date
May 21, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK AS INSTRUCTED. THE USER'S GUIDE STATES TO POWER CYCLE THE SYS AND PERFORM RINSEBACK IF THE ALARM RECURS. THE CYCLER CONTINUES TO BE USED FOR TREATMENTS WITHOUT FURTHER PROBLEMS. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADD'L INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A SYS ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT. THE OPERATOR ENDED TREATMENT WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8017710

Patients

Seq Age Sex Outcome Treatment
1 48 YR Other