AXSYM RUBELLA IGG ANTIBODY
Report
- Report Number
- 1415939-2008-00099
- Event Type
- Malfunction
- Date Received
- June 17, 2008
- Date of Event
- May 6, 2008
- Report Date
- June 2, 2008
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- LFX
- PMA / PMN Number
- K954045
- Removal / Correction Number
- 1415939-6/06/08-004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER CONTACTED ABBOTT REPORTING THEY WERE UNABLE TO CALIBRATE THE AXSYM RUBELLA IGG ASSAY AFTER TWO ATTEMPTS. THE CUSTOMER STATED WEEKLY MAINTENANCE WAS PERFORMED, REPLACED BOTH PROBES AND FLUSHED THE ANALYZER. THE CUSTOMER ATTEMPTED A RECALIBRATION WHICH FAILED. THE CUSTOMER STATED THERE IS NO CALIBRATION ISSUE WITH OTHER AXSYM ASSAYS. THE ABBOTT CUSTOMER TECHNICAL ADVOCATE (CTA) REVIEWED TROUBLESHOOTING PROCEDURES AND CALIBRATION DATA, AND SENT THE CUSTOMER A NEW LOT OF REAGENT. AS PART OF THE TROUBLESHOOTING, THE CTA REQUESTED THE CUSTOMER SPIN RUBELLA CALIBRATOR A AND ATTEMPT RECALIBRATION, WHICH WAS UNSUCCESSFUL. THE CUSTOMER WAS SENT A NEW LOT OF CALIBRATORS FOR TROUBLESHOOTING AND THE CALIBRATOR A VALUES WERE ELEVATED. THE CUSTOMER USED A DIFFERENT LOT OF CALIBRATOR AND WAS ABLE TO ACHIEVE SUCCESSFUL CALIBRATION. THE CTA REQUESTED THE CUSTOMER RETURN THE SUSPECT CALIBRATORS FOR INVESTIGATION. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AXSYM RUBELLA IGG ANTIBODY | MEIA FOR THE MEASUREMENT OF IGG ANTIBODY TO THE RUBELLA VIRUS | LFX | ABBOTT LABORATORIES | NA | 58295M100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Required Intervention | AXSYM RUBELLA CALIBRATORS: LIST#09C29-01| LOT#61236M200| AXSYM RUBELLA CALIBRATORS: LIST#09C29-01| LOT# 62167M200| AXSYM ANALYZER: LIST # 07A83-95| LOT#58184M100 |