FDA Adverse Event Malfunction Summary report: N

AXSYM RUBELLA IGG ANTIBODY

MDR report key: 1061580 · Received June 17, 2008

Report

Report Number
1415939-2008-00094
Event Type
Malfunction
Date Received
June 17, 2008
Date of Event
May 6, 2008
Report Date
June 2, 2008
Manufacturer
ABBOTT LABORATORIES
Product Code
LFX
PMA / PMN Number
K954045
Removal / Correction Number
1415939-6/06/08-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED ABBOTT REPORTING THEY WERE UNABLE TO CALIBRATE THE AXSYM RUBELLA IGG ASSAY TWICE AND OBSERVED ERROR CODE 1018 (CALIBRATION CHECK FAILURE, CALIBRATOR A RESULT TOO HIGH) AND 3057 (AIR DURING ASPIRATION, BUFFER WELL, PROCESSING PROBE). THE CUSTOMER WAS ADVISED TO INSPECT PROCESSING PROBE FOR DAMAGE, AND PERFORM OTHER TROUBLESHOOTING PROCEDURES. AFTER TROUBLESHOOTING, THE CUSTOMER ACHIEVED SUCCESSFUL CALIBRATION. AS THE CUSTOMER OBTAINED FOUR PT RESULTS IN THE "GRAY" ZONE FOR POSITIVE INTERPRETATION, THE CUSTOMER PERFORMED PRECISION STUDY WITH THE AXSYM RUBELLA POSITIVE CONTROL AND A PT SAMPLE. THE FOUR "GRAY" ZONE PT SAMPLES WERE RE-ASSAYED AND VERIFIED AS "GRAY" ZONE, THEREFORE, THE CUSTOMER WAS SATISFIED THE RESULTS WERE VALID AND REPORTED THE RESULTS. THE CUSTOMER RETURNED THE SUSPECT CALIBRATORS FOR INVESTIGATION. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM RUBELLA IGG ANTIBODY MEIA FOR THE MEASUREMENT OF IGM ANTIBODIES TO RUBELLA VIRUS LFX ABBOTT LABORATORIES NA 60943M100

Patients

Seq Age Sex Outcome Treatment
1 NI AXSYM ANALYZER| AXSYM RUBELLA CONTROLS: LOT# 58186M100| AXSYM RUBELLA CALIBRATORS: LOT# 58184M100| AXSYM RUBELLA CALIBRATORS: LOT# 60258M100